Checking »Paw high« alarm
1 Set »V
« knob to 0.5 L.
T
2 Set »Freq.« knob to 12 1/min.
3 Set »T
: T
« knob to 1 : 2.
I
E
4 Set »Pmax« knob to 40 mbar.
Keep test lung compressed and watch pressure
gauge:
At an airway pressure of 36 to 40 mbar, the ventilator
should switch over to expiration and the test lung
deflates.
5 The red alarm indicator flashes and the following
message appears on the display:
Paw high
The intermittent tone sounds.
Release test lung.
The intermittent tone ceases.
g
6 Press »
Reset« to clear the display.
Checking »Paw low« alarm
Use the same settings as above.
Disconnect test lung from breathing valve.
5 The red alarm indicator flashes after approx.
20 seconds and the following message appears on
the display:
Paw low
The intermittent tone sounds.
Reconnect test lung.
The intermittent tone ceases after approx. 25 seconds.
g
6 Press »
Reset« to clear the display.
Checking readiness for operation

Checking "Paw high" alarm

Checking "Paw low" alarm