Source test

Source test is started with menu item "Source test" in submenu
"Settings".
Source test starts with the parameters previously set with rad-
eye.exe. Furthermore it is necessary to set the actual date and
time.
Start the source test without source. The RadEye check the back-
ground value. If the background is lower than 0.15 µSv/h the user
is asked to place the source. High range test is only performed
with RadEye G20-ER and G20-ER10.
4-10