Date and time of last success-
ful source test
Low range test
High range test
After placing the source, source test is started automatically.
During measurement the display shows the actual measurement
value. After measurement the deviation is calculated and dis-
played. The test is successful when the deviation is between -20%
and +20%. During source test the alarm LED is permanently on.
Measured de-
viation
Actual meas-
urement value
RadEye G20
Iw/Tr/Ff 26.11.2012
Terminate
source test with
"►" key
DB-067 E
Start source test
by placing the
source
Test result
Remaining
time
4-11