9. Maintenance

9.1

Recommended maintenance

The RadEye is basically maintenance free.
9.1.1

Source test

It is recommended to perform a source test routine with the
optional Lutetium test adapter on a regular basis. The source
test routine is described in chapter 4.13.
Level of usage
Intensive, changing climate
Normal, daily usage
Low
9.1.2

Cleaning the instrument

The rubber boot should be removed and cleaned separately.
The RadEye unit shall be cleaned with a moist tissue (only
water as cleaner). The rubber boot shall be washed in water.
Do not use aggressive chemicals like organic solvents.
RadEye G20
Iw/Tr/Ff 26.11.2012
Period between Source tests
3 month
6 month
12 month
DB-067 E