8.2

Reading data from the start up screen

The start up screen shows important data like the last calibra-
tion, the current voltage and the firmware version.
This screen stays active as long as the "" button is kept
pressed during the power on of the RadEye device.
8.3
The RadEye doesn't show the menu items as
described in the manual
The items available in the RadEye menu can be disabled or
enabled via the RadEye.EXE software. The factory setting has
all menu functions enabled. Please refer to chapter 7.5, menu
settings. Next, please check the firmware version of your
RadEye (see chapter 2.4) and compare your version number
with the revision table of this manual at the beginning of the
document. The revision table shows the introduction of new
functions together with the according firmware version.
8-2
DB-065 E
Iw/Tr/Ff 06.03.2017