Radeye Inductive Charger - Thermo Scientific RadEye B20 Operating Instructions Manual

Multi-purpose survey meter
Hide thumbs Also See for RadEye B20:
Table of Contents

Advertisement

11.11 RadEye inductive charger

The RadEye holder serves as supporting device and, in combina-
tion with the special battery lid 42506/7034, power supply unit
for the operation of a RadEye instrument in a vehicle. For con-
venient mounting to the dashboard one of the mechanical adap-
tors 42506/7061 ... 64 is required.
Note: The charging function of this device requires the battery
latch 42506/7034. Without the above mentioned part no charging
will be achieved.
Safety instructions
1. Only use NiMH – accumulators, no primary batteries (Alka-
line – manganese, zinc – carbon ...)
2. Only use two accumulators of the same type (manufacturer,
rated capacity ...)
3. Only charge accumulators having the same discharging state
4. Never use defective accumulators (penetrating/emerging elec-
trolyte, damaged housing...)
5. The bottom of the housing (aluminum heat sink) gets warm
(50°C, 120°F max.).
6. Not for use in open convertibles.
7. Do not put cards with magnet stripes (f.e. credit cards, park-
ing cards, phone cards, etc.) near the mounted warning sym-
bol. The magnetic field could erase the data from your card.
8. Clearance of at least 50cm (20") between charger and car
radio is recommended.
RadEye B20/-ER
Iw/Tr/Ff 06.03.2017
DB-065 E
11-13