The scaler mode starts with the parameter set in submenu „Scaler
Parameter".
Preset time
Elapsed time
To start the measurement, press the ▲/Info key. While the meas-
urement is in progress "Stop" is blinking.
The error is calculated as follows:
F: absolute error in cps
RM: measuring value in cps
tM: measuring time in s
R0: Background count rate in cps *)
t0: Background measuring time in s *)
*) Background count rate and Background measuring time are
only used in combination with"Scaler netto".
If measuring unit cpm, Bq, dpm or Bq/cm² is used, the error is
displayed in the corresponding unit. The error is not shown in
dose rate mode or "Accumulated counts" is active.