Display's Top Line - Thermo Scientific RadEye B20 Operating Instructions Manual

Multi-purpose survey meter
Hide thumbs Also See for RadEye B20:
Table of Contents

Advertisement

8.4
The RadEye shows an error message in the

display's top line

Please refer also to chapter 6.2. If one of the following failures
is indicated, then the RadEye must be repaired by the Thermo
Scientific bench repair team:
HV-Error
Detect Err
EEPROM Err
Watchdog*
Please contact our Customer Service for more details.
*) Please see the failure description in the above referenced chapter.
If the failure was related to a strong electromagnetic pulse scenar-
io and the failure went away afterwards, then a repair may not be
necessary. A diligent observation of the RadEye proper operation
after the failure is strongly recommended.
8.5
The RadEye is not found in the software
If the connection between the RadEye and the computer can't
be established with a selected serial port (COM port), then the
program shows the bellow message first:
RadEye B20/-ER
Iw/Tr/Ff 06.03.2017
DB-065 E
8-3

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Radeye b20-er

Table of Contents