SAFETY INSTRUCTIONS
Dose integration is performed only, if the instrument is in a
dose rate measuring mode (R/h; Rem/h Sv/h Gy/h). In case the
measuring range is exceeded, the value of the upper end of the
dose rate measuring range is used for the dose integration.
The RadEye B20 is suited to perform highly accurate dose
measurements. It is however not intended for use as a legal
personal dose meter.
The RadEye B20 is not well suited for pulsed radiation. There
will be a significant under estimation at dose rates during
pulse of more than 0.2 R/h.
Do not use the unit if error messages appear on the screen.
The earphone connector at the bottom side of the instrument
must be exclusively used by equipment that is specified for
use with RadEye B20.
The instrument must not be used in explosive atmosphere.