System description
Technical Data
empty
22
In its original transport packaging, the SIROLaser Advance can withstand
the following ambient transport conditions:
● Temperatures from -40° C to +70° C
● Relative humidity from 10% to 95%
● Atmospheric pressure from 800 hPa to 1060 hPa
Operating conditions
The SIROLaser Advance may be operated in the following environmental
conditions:
● Temperatures from +10° C to +33° C
● Relative humidity from 10% to 95%
● Atmospheric pressure from 800 hPa to 1060 hPa
CAUTION
Following transport and storage, let the SIROLaser Advance adapt to
room temperature for about one hour prior to operation to reduce the risk
of malfunctions caused by condensation.
Sirona Dental Systems GmbH
Operating Instructions SIROLaser Advance
63 74 198 D3545
D3545.201.06.03.02 05.2014