 Installation
Initial start-up – procedure for proper assembly
A
26
4.4.2
Handpiece, single-use tip usage and assembly of optical
fiber

Handpiece

4.4.2.1
The handpiece consists of three parts:
B
A
Stainless steel handpiece back-sleeve
B
Handpiece body with tube
C
Handpiece sleeve

Single-use tip usage

4.4.2.2
1. Please sterilize the single-use tip before use.
2. Attach the single-use tip firmly on the handpiece prior to mounting the
optical fiber.
3. After having mounted the optical fiber bend the single-use tip with the
bending tool to the angle that you need for best handling and
depending on the treatment.
Single use tip, Sirona
WARNING
Do not move the fiber in the single use tip after having bent the tip, risk
of fiber damage!
Always use the tip with the fiber and check for the correct fixation.
If the optical fiber does not protrude at least 1 cm out of the single-use
tip, there is a risk that the single-use tip will heat up.
Do not move the fiber in the tip.
Do not resterilize the single use tips.
To avoid the risk of damaging optical fibers use only Sirona single-use
tips. Also make sure that the single-use tips are only bent with the
bending tool when the optical fiber is already mounted.
CAUTION
Check whether the single-use tip is damaged before attaching it.
Replace it if necessary.
Please check that the single-use tip is firmly attached prior to each use.
Sirona Dental Systems GmbH
Operating Instructions SIROLaser Advance
C
63 74 198 D3545
D3545.201.06.03.02 05.2014