 Installation
Initial start-up – procedure for proper assembly
28
Screwing on the optical fiber
1. Open handpiece by pressing both snap tabs and pull apart the
handpiece sleeve from the handpiece body.
2. Remove the protection screw from the handpiece body by using an
empty transport and sterilization tube.
3. Remove the protective cap from the optical fiber.
4. Insert the optical fiber into the handpiece and screw it on. Tighten the
optical fiber only slightly so that it can be loosened with the transport
and sterilization tube again with ease. Make sure not to overwind the
transport and sterilization tube!
5. Remove the transport and sterilization tube from the optical fiber.
OK
6. Put the laser into operation by choosing any preset treatment. A
corresponding description is provided in chapter "Operation".
The aiming beam must not be aimed toward a person's eye. It
comprises an intensive light source even when set to a low power level.
Always wear protective goggles.
7. As soon as the SIROLaser Advance is ready you can check to make
sure that the aiming beam illuminates evenly, i.e. projects a circular
light pattern (see adjacent picture).To do this, aim the optical fiber
vertically at a white background located approx. 5 cm away. If the
beam shows no pattern or the beam pattern is not illuminated evenly,
the optical fiber may be damaged or defective. In this case, return the
optical fiber to your dental depot so that it can be replaced under
warranty. Do not use any defective optical fibers.
WARNING

Sirona Dental Systems GmbH

Operating Instructions SIROLaser Advance
63 74 198 D3545
D3545.201.06.03.02 05.2014