Base Temperature; Pressure Input; Base Pressure - GE XGS868 Programming Manual

Steam ultrasonic mass flow transmitter
Table of Contents

Advertisement

November 2004

Base Temperature

Pressure Input

Base Pressure

1-12
1. Enter the base temperature and press Next Item/Enter. The ratio of
this value to the actual temperature is used to calculate the
standard mass flow rate.
2. Do one of the following:
If you selected No at Assume Saturation, proceed to Pressure
Input below.
If you selected Pressure as the Steam Input Type, proceed to
Pressure Input below.
If you selected Temperature as the Steam Input Type, proceed
to Base Pressure below.
1. Use the pull-down menu to select a fixed pressure value or to set
up the option card in Slot 1 that will supply the live pressure input
and press Next Item/Enter.
Note: If Slot 1 contains an activated option card with an input
Pressure, Slot 1
assigned to
prompt. If the process pressure is stable, a fixed value may be
used, but most applications require a live pressure input.
If there is no active option card for pressure, the meter
assumes you are using a fixed pressure.
2. Proceed to one of the following sections:
If you selected Fixed - proceed to Step 3.
If you selected Slot 1 - proceed to Step 4.
3. Enter the known fixed process pressure and press Next Item/
Enter. The meter will only accept values from 0 to 5000 psia.
Proceed to Base Pressure below.
4. Use the pull-down menu to set up input A or input B and press
Next Item/Enter. The inputs were labeled during setup.
Note: The set up of input A is used as an example. Identical
procedures would be used to set up input B.
Enter the base pressure and press Next Item/Enter. The ratio of this
value to the actual pressure is used to calculate the standard mass
flow rate.
appears as an option at the above
Programming Site Data

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents