MMY30 Specifications

Sensing Element

Measurement Range

Recommended
Recalibration Cycle

Calibration Accuracy

Maximum Sensor Relative
Humidity
Operating and Storage
Temperature
Air Bleed Off at 7 bar (100
psig)
Maximum Operating
Pressure

Helium Leak-Rate

Output

Flow Block

Wrench Width for Flow
Block

Electronics

Moisture Unit

Power Supply

Protection

Weight

European Compliance

Technical Specifications
Planar sensor, aluminum oxide capacitance principle
- 90°C to +10°C (-130°F to +50°F) dew point temperature.
0 to 10, 0 to 100, 0 to 1000 ppm
display)
12 months, depending on the application
±2°C (±3.6°F) dew point at 25°C (77°F)
50% at dewpoint temperatures >0°C (32 °F)
-40°C to +50°C (-40°F to +122°F)
Approximately 28 sl/h (1 SCFH)
31 bar, 3.1 MPa (450 psig)
-6
<10
mbar l/s
4 to 20 mA; 16 µA resolution
316 stainless steel with G ½ thread (DIN ISO 228) or ½" (12.7 mm)
MNPT thread
42 mm (1 5/8")
Microcontroller operated
Dew point temperature in °C or °F, ppm
24 VDC nominal, 12 to 28 VDC range
Type 4X (IP 67)
1.5 kg (3.2 lbs)
Complies with EMC Directive 89/336/EEC and PED 97/23/EC for
DN<25
November 2005
(fully adjustable with integral
v
v
6-1

Hide quick links: