Pipe Data - GE XGS868 Programming Manual

Steam ultrasonic mass flow transmitter
Table of Contents

Advertisement

November 2004

Pipe Data

1-8
If a standard transducer is being used, the programming sequence
should be rejoined at this point.
1. Enter the known pipe outside diameter or circumference, then use
the pull-down menu to select the units. Press Next Item/Enter.
Obtain the required information by measuring either the pipe
outside diameter (OD) or circumference at the transducer
installation site. The data may also be obtained from standard pipe
size tables found in Sound Speeds and Pipe Size Data manual
(914-004). Available units are shown in Table 1-3 below.
Table 1-3: Available Pipe OD Units
English
pipe OD in inches
pipe OD in feet
pipe circumference in inches
pipe circumference in feet
2. Enter the know thickness of the pipe wall and press Next Item/
Enter. If the pipe wall thickness is not available, look up the value
in a table of standard pipe size data which can be found in Sound
Speeds and Pipe Size Data manual (914-004).
IMPORTANT:
Because the units cannot be independently chosen for
this parameter, the value must be entered in the same
units used for the pipe OD.
3. Enter the path length of the ultrasonic signal, then use the pull-
down menu to select the units. Press Next Item/Enter.
Note: If a spoolpiece was ordered with the meter, the transducer
signal path length
are engraved on the flowcell and/or are included in the
documentation supplied with the meter. For on-site transducer
installations, refer to Appendix C, Measuring P and L
Dimensions, in the Startup Guide for instructions.
Metric
pipe OD in millimeters
pipe OD in meters
pipe circumference in millimeters
pipe circumference in meters
(P)
and the transducer signal axial length
Programming Site Data
(L)

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents