Procedure Options
Programming Site Data
Setting Up the Totalizer
1. Use the pull-down menu to select the desired channel option and
press Next Item/Enter.
2. Use the pull-down menu to select the desired measurement
parameter and press Next Item/Enter. See Table 1-11 below for a
description of the available options.
Table 1-11: Output Measurement Options
Forward Totalized Volume Flow
Reverse Totalized Volume Flow
Forward Totalized Mass Flow
Reverse Totalized Mass Flow
Note: The measurement units that appear in these prompts are those
Global-System
selected in the
3. Enter a value for the minimum pulse on time (between 1 µsec and
10,000 µsec) for the frequency of the totalizer pulses and press
Next Item/Enter.
Note: A complete pulse consists of equal amounts of ON and OFF
times. Choose a value that is compatible with the frequency
counter to be used.
4. Enter a value for the number of measurement units represented by
each pulse and press Next Item/Enter.
After completing the above steps, IDM returns to the Input/Output
window. Do one of the following:
•
To continue programming, refer to Appendix A, Menu Maps
for IDM, to navigate to the desired menu.
To leave the User Program, press Exit Page three times.
•
November 2004
Options
window earlier in this section.
1-33