GE XGS868 Programming Manual page 41

Steam ultrasonic mass flow transmitter
Table of Contents

Advertisement

November 2004
Frequency Outputs
1-34
The frequency output issues a continuous signal with a frequency
proportional to the selected measurement. While following the
programming instructions, refer to Figure A-4 in Appendix A.
Complete the following steps to set up the totalizer outputs of an
SLOT 1
option card installed in
Accessing the Frequency Output
1. Open the SITE EDIT MENU from the Edit Functions menu.
2. Select Global.
3. Select Input/Output.
4. Select Options.
5. Select Slot 1.
6. Use the pull-down menu to select the desired input and press Next
Item/Enter.
7. Do one of the following:
If you selected Off, IDM returns to the Input/Output window.
Go to Procedure Options on the following page.
If you selected Frequency, and are using a
1-Channel meter, proceed to Step 2 in the next section.
2-Channel meter, proceed to Step 1 in the next section.
Note: Output letters (A, B, etc.) correspond to the output wiring
terminal numbers from top to bottom. For example, Output A
corresponds to pins 1, 2 and 3; Output B corresponds to pins
4, 5 and 6; etc.
Setting Up the Frequency
1. Use the pull-down menu to select the desired channel option and
press Next Item/Enter.
2. Use the pull-down menu to select the desired measurement
parameter and press Next Item/Enter. See Table 1-12 on the next
page for a description of the available options. Refer to page 3-2 in
the Service Manual for a description of diagnostic parameters.
: