8.1.4
Setting Output Smoothing Coefficient
When the oxygen concentration in the sample gas fluctuates rapidly, if the measured value is
used for control this can lead to problems with undesirable frequent ON/OFF switching. You can
set a smoothing time constant of between 0 and 255 seconds to reduce the effect. Select the
appropriate output damping constant from the numeric-data entry display. To set 30 seconds,
enter 030.
8.1.5
Selection of Output Mode
You can select whether the relationship between the sample oxygen concentration and the
analog output signal be linear or logarithmic. Press the [Enter] key in the output mode display. A
linear/ logarithmic selection display then appears. Select the desired mode.
NOTE
If you select an output mode of "logarithmic" then regardless of range setting the minimum output
value becomes 0.1 vol%O
8.1.6

Default Values

When the analyzer is delivered or reset to defaults, the output current default settings by as
shown in Table 8.1.
Table 8.1
Item
Min. oxygen concentration
Max. oxygen concentration
Minimum humidity conc.
Maximum humidity conc.
Minimum mixing ratio
Maximum mixing ratio
Output damping constant
Output mode
8.2

Output Hold Setting

The "output hold" functions hold an analog output signal at a preset value during the equipment's
warm-up time or calibration or if an error arises. Outputs 1 and 2 can not be set individually. Table
8.2 shows the analog outputs that can be retained and the individual states.
Table 8.2
Equipment status
Output
hold values available
4 mA
20 mA
Without hold feature
Retains output from just before
occurrence
Set value (2.4 to 21.6 mA)
: The output hold functions are available.
*1: The output hold functions on error occurrence are unavailable when option code "/C2" or "/C3"(NAMER NE 43 compliant) is
specified.
(humidity; 0.1% H
2
Output Current Default Values
Default setting
0 vol%O
2
25 vol%O
2
0 % H
O
2
25 % H
O
2
0.000 kg/kg
0.200 kg/kg
0 (seconds)
Linear
During warm-up
<8. Detailed Data Setting>
O, mixing ratio; 0.01 kg/kg) fixed.
2
Under
Under calibration
maintenance
During blow back
IM 11M12A01-02E
8-4
On Error
occurrence(*1)
16th Edition : Mar. 19, 2020-00