Mounting Of Cable Gland; Wiring - YOKOGAWA ZR22G User Manual

Separate type zirconia oxygen/humidity analyzer
Hide thumbs Also See for ZR22G:
Table of Contents

Advertisement

Table 5.1
Terminal name of converter
CELL+, CELL-
HTR TC+, HTR TC-
CJ+, CJ-
HEATER
L, N
AO-1+, AO-1-, AO-2+, AO-2-
DO-1, DO-2, DO-3, DO-4
AC-Z, AC-S, AC-C
DI-1, DI-2, DI-C
AI+, AI-
Note *:
When the case is used for protective grounding, use a 2-core cable.
WARNING
Cables that withstand temperatures of at least 80°C should be used for wiring.
CAUTION
• Select suitable cable O.D. to match the cable gland size.
• Protective grounding should be connected in ways equivalent to JIS D style (Class 3)
grounding (the grounding resistance is 100 Ω or less).
• Special consideration of cable length should be taken for the HART communication, For the
detail, refer to Section 1.1.2 of the IM 11M12A01-51E "Communication Line Requirements".
5.1.1
Terminals for the External Wiring in the Converter
Open the front door and remove the terminal covering plate to gain access to the converter
external wiring terminals (see Figure 5.1).
CAUTION
After wiring necessary cable to the converter terminals, be sure to fix the terminal covering plate
with two screws again.
Cable specifications
Name
Detector signal
Detector heater
Power supply
Analog output
Contact output
Automatic Calibration unit
Contact input
Temperature input
<5. Wiring>
Need for shields
Number of cores
О
2 or 3 *
О
2 or 4
2 to 8
О
IM 11M12A01-02E
5-2
6
2
3
3
2
16th Edition : Mar. 19, 2020-00