Setting Calibration Time - YOKOGAWA ZR22G User Manual

Separate type zirconia oxygen/humidity analyzer
Hide thumbs Also See for ZR22G:
Table of Contents

Advertisement

(2) Record of zero correction ratio
Recorded the past ten zero correction ratios.
(3) Response time
You can monitor the response time provided that a two-point calibration has been done in semi-
automatic or automatic calibration.
(4) Cell's internal resistance
The cell's internal resistance gradually increases as the cell (sensor) deteriorates. You can
monitor the values measured during the latest calibration. However, these values include the
cell's internal resistance and other wiring connection resistance. So, the cell's degrading cannot
be estimated from these values only.
When only a span calibration has been made, these values will not be measured, and previously
measured values will remain.
(5) Robustness of a cell
The robustness of a cell is an index for predicting the remaining life of a sensor and is expressed
in a number on four levels.
9.2
Calibration Procedures
NOTE
Calibration should be made under normal operating conditions (if the probe is connected to a
furnace, the analyzer will undergo calibration under the operating conditions of the furnace).
To make a precise calibration, conduct both zero and span calibrations.
The following sets forth the required calibration settings:
9.2.1
Mode
There are three calibration modes available:
(1) Manual calibration which allows zero and span calibrations or either one manually in turn;
(2) Semi-automatic calibration which lets calibration start with the touchpanel or a contact input,
and undergoes a series of calibration operations following preset calibration interval and
stabilization time.
(3) Automatic calibration which is carried out automatically following preset calibration interval.
Calibrations are limited by the following mode selection:
• When Manual calibration is selected:
Manual calibration only can be conducted. (This mode does not allow semi-automatic calibration
with a contact input nor automatic calibration even when its start-up time has reached.)
• When Semi-automatic calibration is selected:
This mode enables manual and semi-automatic calibrations to be conducted. (The mode,
however, does not allow automatic calibration even when its start-up time has reached.)
• When Automatic calibration is selected:
This calibration can be conducted in any mode.
To execute this calibration, follow these steps:
(1) Select the [Setup] key from the Basic panel display to display the Execution/Setup display.
Then select "Maintenance" from the Execution/Setup display.
<9. Calibration>
IM 11M12A01-02E
16th Edition : Mar. 19, 2020-00
9-6