A pipe to be mounted
(nominal JIS 50A : O.D. 60.5 mm)
Figure 3.20
3.7
Insulation Resistance Test
Even if the testing voltage is not so great that it causes dielectric breakdown, testing may cause
deterioration in insulation and a possible safety hazard. Therefore, conduct this test only when it
is necessary.
The applied voltage for this test shall be 500 V DC or less. The voltage shall be applied for as
short a time as practicable to confirm that insulation resistance is 20 MΩ or more.
Remove wiring from the converter and the detector.
(1) Remove the jumper plate located between terminal G and the protective grounding terminal.
(2) Connect crossover wiring between L and N.
(3) Connect an insulation resistance tester (with its power OFF). Connect (+) terminal to the
crossover wiring, and (-) terminal to ground.
(4) Turn the insulation resistance tester ON and measure the insulation resistance.
(5) After testing, remove the tester and connect a 100 kΩ resistance between the crossover
wiring and ground, to discharge for over 1 min. During discharge, do not touch the terminal.
(6) Testing between the heater terminal and ground, contact output terminal and ground, analog
output/input terminal and ground can be conducted in the same manner.
(7) Although contact input terminals are isolated, insulation resistance test cannot be conducted
because the breakdown voltage of the surge-preventing arrester between the terminal and
ground is low.
(8) After conducting all the tests, replace the jumper plate as it was.
F3-16E.ai
Pipe Mounting
<3. Installation>
IM 11M12A01-02E
16th Edition : Mar. 19, 2020-00
3-13