Installation Of Zr40H Automatic Calibration Unit - YOKOGAWA ZR22G User Manual

Separate type zirconia oxygen/humidity analyzer
Hide thumbs Also See for ZR22G:
Table of Contents

Advertisement

126.5
190
Figure 3.10
Mounting holes
(2) Mount the converter. Secure the converter on the wall using four screws.
Note: For wall mounting, the bracket and bolts are not used.
Figure 3.11
Wall Mounting
<Panel Mounting>
(1) Cut out the panel according to Figure 3.12.
+2
183
0
Figure 3.12
Panel cutout sizes
(2) Remove the fitting from the converter by loosening the four screws.
(3) Insert the converter case into the cutout hole of the panel.
(4) Attach the mounting fitting which is once removed in step (2) again to the converter.
(5) Firmly fix the converter to the panel. Fully tighten the two clamp screws to hold the panel
with the fitting.
Unit: mm
Four holes 6 mm in diameter
for M5 screws
F3-8E.ai
F3-9E.ai
Unit: mm
+2
274
0
F3-10E.ai
Figure 3.13
<3. Installation>
Clamp screw
Panel mounting
IM 11M12A01-02E
3-9
Washer
Bolt
Fitting
F3-11E.ai