Confirmation of Converter Type Setting
This converter can be used for both the Oxygen Analyzer and the Humidity Analyzer.
Before setting the operating data, be sure to check that the desired converter model has been
set.
Note that if the converter type setting is changed, the operating data that have been set are then
initialized and the default settings remain. To set the desired operating data, follow these steps:
(1) Press the [Setup] key.
(2) Use the [▼] key to select "Commissioning" and press the [Enter] key.
(3) In the password display, enter the [Enter] key. If the password is to be set again, enter the
new password (for details, see Section 8.6.6, Setting Passwords, later in this manual).
(4) The Commissioning display shown in Figure 7.4 appears. Select "Basic setup" and press
the [Enter] key.
(5) The Basic setup display shown in Figure 7.5 then appears. Confirm the currently set
converter type. If the Humidity Analyzer option /HS was selected at the time of purchase, the
converter was set for high temperature humidity use before shipment.
(6) If the converter type is to be changed, press the [Enter] key. The display shown in Figure 7.6
then appears.
(7) Use the [▼] key to select the type of equipment. Then press the [Enter] key to complete the
converter selection.
(8) If the type of converter is changed after setting the operating data, those data are then
initialized and the default settings remain. Reset the operating data to meet the new type of
equipment.
Commissioning
Basic setup
mA-output setup
Alarms setup
Contact setup
Others
Figure 7.4
Basic setup
This instrument is for
Oxygen
Humidity
Please note:changing this
Basic setting will return ALL
other setting to Default!!
Detector is ZR22
Select measure gas: Wet
Figure 7.6
Enter
F7.6E.ai
Commissioning Display
Enter
Equipment Setup
Basic setup
This instrument is for
Oxygen
Please note:changing this
Basic setting will return ALL
other setting to Default!!
Detector is ZR22
Select measure gas: Wet
Figure 7.5
F7.8E.ai
<7. Startup>
Enter
F7.7E.ai
Basic Setup
IM 11M12A01-02E
16th Edition : Mar. 19, 2020-00
7-3