9.3.2
Semi-automatic Calibration
To start calibration, follow these steps:
(1) Press the [Setup] key in the Basic panel display to display the Execution/Setup display.
Then select "Calibration" from the Execution/Setup display. The Calibration display shown in
Figure 9.11 appears.
(2) Select "Semi-auto calibration" to display the Semi-auto calibration display shown in Figure 9.12.
(3) Select "Start calibration". The display shown in Figure 9.13 appears, and then start calibration.
Calibration
Manual calibration
 Semi-auto calibration
Figure 9.11
Semi-auto calibration
2 1 . 0 0%
1 . 0 0%
2 1 . 0 % O2
Figure 9.13
To start calibration using an input contact, follow these steps:
(1) Make sure that Calibration start has been selected in the Input contacts display (see Section
8.5, earlier in this manual).
(2) Apply an input contact to start calibration.
To stop calibration midway, follow these steps:
(1) Press the [Reject] key. If this key is pressed midway during calibration, the calibration will
stop and the hold (output stabilization) time will be set up.
(2) Press the [Reject] key once again to return to the Basic panel display and the analyzer will
be in normal measurement.
9.3.3
Automatic Calibration
No execution operations are required for automatic calibration. Automatic calibration starts in
accordance with a preset start day and time. Calibration is then executed at preset intervals.
NOTE
Before starting Semi-automatic calibration or Automatic calibration, operate the calibration gas
solenoid valves and adjust calibration gas flow to 600 ± 60 ml/min.
Enter
Calibration Display
0.5min./div
CAL. TIME
Semi-auto Calibration Display
Semi-auto calibration
Check settings
Zero gas conc:
Span gas conc: 2 1 . 0 0 %
Cal time: 1 0 min 0 0 s
Go to setup to change value.
Start calibration
 Cancel calibration
F9.8E.ai
Figure 9.12
F9.10E.ai
<9. Calibration>
1 . 0 0 %
Enter
Semi-auto Calibration Display
IM 11M12A01-02E
16th Edition : Mar. 19, 2020-00
9-10
F9.9E.ai