Parts Required For Piping In System 1 - YOKOGAWA ZR22G User Manual

Separate type zirconia oxygen/humidity analyzer
Hide thumbs Also See for ZR22G:
Table of Contents

Advertisement

Insulation
resistance tester
-
+
crossover wiring
1
2
3
4
FG
AO1
AO1
AO2
(+)
(-)
(+)
12
13
14
15
FG
DI-1
DI-2
DI-C
Contact
input 2
Contact
input 1
23
24
25
26
DO-1
DO-1
DO-2
DO-2
DO-3
Remove wiring to connect external circuits
and Connect crossover wiring
Insulation
resistance tester
-
+
Figure 3.21
ZR402G Converter
5
6
7
8
9
10
AO2
CJ
CJ
TC
TC
CELL
(-)
(+)
(-)
(+)
(-)
(+)
16
17
18
19
20
21
AI
AI
AC-
AC-
AC-
FG
(+)
(-)
ZERO
SPAN
COM
Solenoid valve
Span gas
for automatic
calibration
Solenoid valve for
automatic calibration
Zero gas
Automatic Calibration Unit
27
28
29
30
31
32
DO-3
DO-4
DO-4
HTR
HTR
Insulation
resistance tester
-
<3. Installation>
11
CELL
CELL
CELL
(-)
(+)
(-)
22
FG
33
34
35
36
L
N
G
Remove
jumper plate
crossover wiring
+
IM 11M12A01-02E
3-14
ZR22G Detector
TC
TC
CJ
CJ
(+)
(-)
(+)
(-)
HTR HTR
F3-17E.ai
16th Edition : Mar. 19, 2020-00

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Zr402gZo21rZo21pZh21bZa8fZr40h ... Show all

Table of Contents