Step
3
4
5
6
7
8
9
10
ÄKTA pure User Manual 29119969 AB
Action
Fill a syringe with distilled water, and connect the syringe to the Luer connector.
Squirt distilled water through the Conductivity flow cell in small amounts. Disconnect the
syringe.
Fill a syringe with 1 M NaOH, and connect the syringe to the Luer connector.
Squirt 1 M NaOH through the Conductivity flow cell about five times.
Leave the liquid in the flow cell for at least 15 minutes.
Fill a syringe with distilled water. Connect the syringe to the Luer connector.
Inject the distilled water into the Conductivity flow cell to rinse the flow cell. Disconnect
the syringe.
Disconnect the Luer connector from the top of the Conductivity flow cell, and reconnect
the fingertight connector with tubing.
7 Maintenance
7.6 Maintenance when required
7.6.10 Clean the Conductivity flow cell