9 Reference information

9.4 Recommended tubing kits for prepacked columns

9.4 Recommended tubing kits for prepacked columns
This section specifies recommended tubing kits for prepacked columns used with ÄKTA pure.
Note:
Column name
All HiTrap 1 ml
(including HisTrap™, GSTrap™,
StrepTrap™)
All HiTrap 5 ml
(including HisTrap, GSTrap except
4B, StrepTrap)
GSTrap 4B, 5 ml
All HiScreen
RESOURCE™ Q, S, ETH, ISO, PHE and
RPC 1 ml
RESOURCE RPC 3 ml
RESOURCE Q, S 6 ml
SOURCE™ 15RPC ST 4.6/100
SOURCE 15Q. S, PHE 4.6/100 PE
Mini Q™, Mini S™ 4.6/50 PE
428
Where alternative tubing kits are listed for a given column, the Pressure control option may
be used instead of changing the tubing kit to adapt to different run conditions. Refer to Sec-
tion 9.15 Pressure control, on page 501 for details.
Recommended Tubing Kits
Standard sys-
1
tem
0.5
0.5 or 0.75
0.5 or 0.75
0.5 or 0.75
0.5
0.5
0.5 or 0.75
0.5
0.5
0.25 or 0.5
2
Advanced system
0.5 with Pressure
control or 0.75
0.75
0.5 with Pressure
controlor 0.75
0.5 with Pressure
control or 0.75
0.5
0.5
0.75
0.5
0.5
0.25 or 0.5
Comments
0.75 with Pressure control for
maximum flow rate combined
with high viscosity.
0.75 with Pressure control for
maximum flow rate combined
with high viscosity.
Avoid combination of column
valve V9-C and pH position in
valve V9-pH.
Resolution may be improved with
0.25.
Avoid combination of column
valve V9-C and pH position in
valve V9-pH.
Resolution may be improved with
0.25.
Use 0.25 for maximum resolution.
Avoid column valve V9-C and pH
position in V9-pH.