5 Operation
5.8 Fractionation

5.8.2 Prepare Fraction collector F9-R

5.8.2
Prepare Fraction collector F9-R
Introduction
This chapter describes how to prepare and assemble Fraction collector F9-R before a run.
Fraction collector F9-R is connected to ÄKTA pure and controlled by UNICORN. Control of the fraction
collector can be achieved automatically in a method run, or manually.
Prepare the fraction collector
Before starting to prepare the Fraction collector F9-R, check the fractionation settings in the method
to be run. Perform the steps described below according to the settings in the method.
• Assemble the Tube rack
• Insert collection tubes
• Adjust the Delivery arm
• Update System Settings in UNICORN
Assemble the Tube rack
Illustration of the Tube rack
Note:
200
1
6
5
3
4
Note that the tube guide has both single and L-shaped cutouts, while the tube holder only has
single cutouts.
Part
1
2
2
3
4
5
6
Function
Single cutout
L-shaped cutout
Bowl
Tube support
Tube guide
Tube holder