Service and maintenance
6.1
Safety notes
The following safety note applies to the barrier box and the optics housings.
Impermissible repair of explosion protected devices
Risk of explosion in hazardous areas
• Repair must be carried out by Siemens authorized personnel only.
Commissioning and operation with pending error
If an error message appears, correct operation in the process is no longer guaranteed.
• Check the gravity of the error.
• Correct the error.
• If the error still exists:
Electrostatic charge
Risk of explosion in hazardous areas if electrostatic charges develop, for example, when
cleaning plastic surfaces with a dry cloth.
• Prevent electrostatic charging in hazardous areas.
Improper connection after maintenance
Risk of explosion in areas subject to explosion hazard.
• Connect the device correctly after maintenance.
• Close the device after maintenance work.
Refer to Installing/mounting (Page 21).
LDS 6 & Optics housing CD 6 Compact Operating Instructions
Compact Operating Instructions, 12/2022, A5E03314905-AB
WARNING
WARNING
– Take the device out of operation.
– Prevent renewed commissioning.
WARNING
WARNING
6
51