Statistics

Combination options:
D
78
Cubis MSU User Manual
Statistics
Purpose:
Saving and statistical analysis of weight values and other calculated values.
The values are generated as results:
– Transaction counter
– Sum of all values
– Average value
– Standard deviation
– Lowest value (minimum)
– Highest value (maximum)
– Difference between maximum and minimum
Options:
You can save a maximum of 65,535 components.
Density determination, calculation, averaging, weighing in percent, parts counting,
timer-controlled functions, checkweighing, mass unit conversion, second tare
memory, individual identifiers (SQmin function, DKD measurement uncertainty)
Configuring a Task: Statistics
* = Factory setting
Define how the weight value will be applied.
t
*
The weight value is applied as soon as the corresponding button
Manually:
is touched.
Automatically: first value at stability:
matically as soon as the balance has stabilized after filling and the value exceeds
the minimum load.
Automatically: last value at stability
is applied automatically, as soon as the balance is unloaded and the value falls
short of the minimum load. If stability was reached repeatedly between measure-
ments, the last stable weight value will be stored.
You must now select the minimum load to be applied:
(
none, 10, 20, ...*100 ... 1000 digits
Weighing Mode with Automatic Storage:
Select the weighing mode to be used for the weight values.
t
Initial weighing of samples in a container.
:
Final readout
: The last weight value with stability
).
Final weighing of samples in a container.
The weight value is applied auto-
Weighing in
: