Statistics

N
Purpose
With this application, you can have
weight values and calculated values
totalized and statistically evaluated.
The values determined for the
evaluation are:
– average (mean value)
– standard deviation
– variation coefficient
– sum of all values
– lowest value (minimum)
– highest value (maximum)
– difference between the minimum and
the maximum
You can use the statistics application in
combination with any program chosen
from Application 1 (such as counting,
weighing in percent) and one from
Application 2 (checkweighing, time-
controlled functions) as well as with the
extra functions.
100
Features
– Storage of weight values and calculated
values
– Simultaneous storage of net and
calculated values
– Optional configuration in the Setup
menu for loading weight values and
calculated values either from Applica-
tion 1 (such as counting, weighing in
percent) or from Application 2 (check-
weighing, time-controlled functions)
– Totalizing memory for up to 65,535
values
– Simultaneous display in the text line of
the transaction counter and, e.g., the
current total
– Optional configuration in the Setup
menu for having the balance tare auto-
matically after a value has been stored
in the totalizing memory
– Manual input of the number of
individual weighing operations and
nDef
confirmation using the
(target no. of operations nDef). Result
printed and memory cleared after
printout of nDef
– Optional configuration in the Setup
menu to add the current weight, with
display accuracy, to the current total by
M+
pressing the
soft key and generate a
printout of the result
– Optional configuration in the Setup
menu for stability-dependent storage
of the measured value: Balance/scale
functions, Stability range
– Optional automatic storage of measured
values
Storage of measured value is indicated
by
;
Uu
indicates that you can place a load
zz
on the balance
– Minimum load threshold for automatic
storage
M-
– Press the
value added to the totalizing memory.
The transaction counter value is
reduced by one and a printout is
generated
MR
– Press the
about number of transactions and the
current total. By configuring the Setup
menu, you can define whether the
information is displayed and printed, or
only printed, and whether the informa-
tion comprises an intermediate
or final evaluation
– In the Info window you can use the
o
(
) soft keys to choose which value
l
will be displayed in the text line during
weighing
– Printout of the final result depending
on the Application 1 or Application 2
parameters. Configure the Setup menu
to define which values are included on
the printout (printout of individual
components)
MR
– Press
for a printout of an inter-
soft key
mediate evaluation after each addition
or a final evaluation
– A final evaluation is printed when the
statistics routine is ended by pressing
c, if no final evaluation was gener-
ated by pressing
– Optional configuration in the Setup
menu to clear the statistics memory and
reset the transaction counter by press-
ing c or after an evaluation is print-
ed out
– Totalization data and transaction
counter data are stored in the non-
volatile memory
– Continue totalization after turning the
balance off and back on
soft key to delete the last
: soft key for information
,
q
MR