Running The "Statistics" Application; Running The Sqmin Minimum Initial Weight Application - Sartorius Entris II BCA Advanced Series Operating Instructions Manual

Table of Contents

Advertisement

Available languages
  • EN

Available languages

  • ENGLISH, page 3
Operation
Procedure
1
8 .7 .2

Running the "Statistics" Application

The following values are saved and analyzed:
− Number of components
− Mean value
− Standard deviation
− Variation coefficient sum of all values
− Lowest value (minimum)
− Highest value (maximum)
− Spread: Difference between maximum and minimum
Requirements
− A printer or a PC is connected and configured .
− The "Statistics" application is selected .
Procedure
Place a sample on the weighing pan .
t
To save the weight value: Press the [Apply] button .
t
The position of the saved value is displayed, e .g .,
y
"Component 1" .
The recording of the statistics starts .
y
Remove the sample being weighed .
t
To save the next value: Place a new sample on the
t
weighing pan and press the [Apply] button .
To display the statistical analysis: Press the [Report]
t
button .
To print the statistical analysis: Press the [Print]
t
button .
To exit the application and delete the saved values:
t
Press the [End] button .
40
Entris
II Advanced Line Operating Instructions
®
Press the [Toggle be-
t
tween weight units]
button (1) .
Set the unit and the
t
accuracy .
Confirm with the
t
[Confirm] button .
8 .7 .3
Running the SQmin Minimum Initial
Weight Application
The SQmin application compares the current weight
value with a defined minimum initial weight . The defined
minimum initial weight is defined in a quality assurance
system, for example . This ensures that the weight values
are above the defined minimum initial weights and thus
meet the requirements of the United States
Pharmacopeia (USP)
SQmin is not the same as the minimum load
"Min" in legal metrology .
Requirements
− The SQmin application must only be set up and
activated by Sartorius Service.
− The "SQmin" application is activated .
Procedure
Place a sample on the weighing pan .
t
If a red SQmin value is displayed: The weight value is
y
less than the defined minimum initial weight . The
weight value is displayed in gray . The weight value
cannot be saved and cannot be printed out in
applications (depending on the setting in the "Safety
level" menu) .

Advertisement

Table of Contents
loading

Table of Contents