Using the Balance
Verified balances feature the
following additional displays:
In the line for metrological data:
Min Minimum weight
e Verification scale interval
In the value range:
o calculated values
k Values < 0 and other values
pcs Piece count
32
Cubis MSU User Manual
Display
The display varies depending on whether the application software is in operating
mode or a menu is currently open (setup mode).

Display in Operating Mode

1
2
3
4
5
6
1
Function display: Current task (here: step 1 of 2)
2
Metrology line: Left: weighing capacity (max); right: d= readability (resolution)
of the balance
Value range with current measurement value (the numeric value is displayed
3
in black numbers once the weighing system reaches stability; beforehand it
is shown in gray)
4
Status and warning messages (see below)
5
Task area with instructions on how to proceed
6
Functions which are currently available
User field: Display of current user, date, and time
7
8
For _.CE model balances verified for use as legal measuring instruments:
Display of weights that are not weight values, e.g., calculated values
and values < 0
9
Scaled display of measured value (percentage of weighing capacity utilization)
Status and Warning Messages
The following messages may be shown here:
isoCAL: isoCAL is active (extended temperature range)
isoCAL flashes: Calibration required
Level me! flashes: Balance is not level; leveling required
GLP: GLP print is active; header will be printed
SQmin: The entered SQmin value is permanently displayed
SQmin flashes: The weight value is less than the entered SQmin value
U or U* or PA: The selected DKD value (measurement uncertainty)
is currently displayed.
Ion: Ionizer in progress (only if your model has an ionizer)