Averaging

Purpose:
This application is used for moving samples (e. g., live animals) and for weighing in
unstable environments. A measurement cycle is automatically carried out with a
defined number of measurements for each object to be weighed. The individual
measurements are averaged, and this average is displayed as the result.
For balances used in legal metrology, follow the directions in Section "Executing
Tasks."
Options:
You can start the measurement manually or it can start automatically as soon as
the min. weight is reached. Depending on how strong the unsteadiness of the
sample is, you can vary the sensitivity for the measurement. The number of
measurements from which the average is taken can also be defined.
Combination options:
Checkweighing, timer-controlled functions, totalizing, statistics, formulation,
SQmin function, DKD measurement uncertainty, second tare, individual identifiers
Configuring a Task: Averaging
D
t
Start by pressing
*Automatic
min. load is exceeded.
t
none, 10, 20, 50, *100, ...1000 digits
The number of digits depends on the balance resolution (see Display).
Example: For a balance with a resolution of d = 0.1 (100 mg), 100 digits means
a min. load of 100 + 0.1 = 10 g. With this setting, the automatic start would
be triggered as soon as a min. of 10 g was placed on the balance.
Automatic start only:
t
Select how the application should start.
This starts the application manually.
key:
The application starts as soon as a pre-defined
start:
Define the min. load for the automatic start of the averaging and/or for
the release of the result display for the manual start.
Determine the weighing mode: *
or
Weighing in
Final readout
Cubis MSU User Manual
Averaging
85