In this example, calibration is shown for a Controller with thermocouple/infrared temperature sensor set
as the input type.
adj
30
• Input types 5, 7, 11, 12,
15, 19, 20:
t 54
69as
• Input types 6, 8, 9, 10, 13,
14, 16, 17, 18, 21, 22, 23,
24:
t 24
e220
t -6
2988
E5@C-T Digital Temperature Controllers User's Manual (H185)
1. Connect the power supply.
2. Connect a standard DC current/voltage source (STV), precision digital multimeter
(DMM), and contact junction compensator (e.g., a zero controller as in the figure)
to the thermocouple input terminals, as shown in the figure below.
STV
DMM
Leave open.
Zero
controller
OUTPUT
INPUT
3. Turn the power ON.
4. Move to the Calibration Level.
This starts the 30-minute aging timer. This timer provides an approximate timer for
aging. After 30 minutes have elapsed, the No. 2 display changes to 0. You can
advance to the next step in this procedure even if 0 is not displayed.
5. When the M Key is pressed, the status changes as shown to the left.
The No. 2 display at this time shows the currently entered count value in
hexadecimal. Set the STV as follows:
• Input types 5, 7, 11, 12, 15, 19, 20: Set to 54 mV.
• Input types 6, 8, 9, 10, 13, 14, 16, 17, 18, 21, 22, 23, 24: Set to 24 mV.
Allow the count value on the No. 2 display to fully stabilize, then press the D Key to
temporarily register the calibration settings.
If this count value is outside of the specified range, the No. 2 display will flash and
the count value will not be temporarily registered.
6. When the M Key is pressed, the status changes as shown to the left.
Set the STV to −6 mV.
Allow the count value on the No. 2 display to fully stabilize, then press the D Key
to temporarily register the calibration settings.
If this count value is outside of the specified range, the No. 2 display will flash and
the count value will not be temporarily registered.
7 User Calibration
+
Compensating conductor of currently selected
thermocouple.
Use K thermocouple compensating conductor
for E, R, S, B, W, and PLII thermocouples and
for an infrared temperature sensor.
7
7 - 5