N5172B/82B Test Procedure - Keysight N5171B EXG Service Manual

X-series signal generators
Hide thumbs Also See for N5171B EXG:
Table of Contents

Advertisement

Post-Repair Procedures and Operational Performance Verification
Leveled Output Power

N5172B/82B Test Procedure

Test Setup
1. Connect the equipment as shown:
2. Zero and calibrate the power sensor.
USB U2000A Series Power Sensors do not require the sensor to be zeroed or calibrated.
3. Connect the power sensor to the RF output of the signal generator. Once
4. Preset the signal generator: Press Preset
5. Turn the internal power meter channel on: Press Aux Fctn > Power Meter
6. Turn RF on: Press RF On/Off so that the RF On/Off LED lights.
7. Turn modulation off: Press Mod On/Off so that the Mod On/Off LED turns
8. Set the frequency: Press Frequency and enter the first frequency value
9. Set the amplitude: Press Amplitude and enter the amplitude value for that
10.Measure the output power level.
11.Repeat steps 8 through 10 to measure power at each of the frequencies
12.Confirm that the measured power levels are within the limits listed in the
Without
Modulation
Keysight X-Series Signal Generators Service Guide
the LED on the power sensor turns off, the sensor is ready to measure.
Measurements > Channel A On (or Channel B On).
off.
listed in
Table
4-7.
frequency.
listed in
Table
4-7.
table.
a.
Set the signal generator frequency to the first value listed in
4-7:
Press Frequency > 50 > MHz.
b. Set the amplitude to 7 dBm:
Press Amplitude > 7 > dBm.
Table
255