Assembly Replacement

Inner Bottom Cover

Inner Bottom Cover

Tools Required

— T-10 driver

Removal Procedure

Refer to
1. Disconnect the power cord.
2. Remove the outer-cover from the signal generator. Refer to
3. Place the signal generator flat with the front-panel facing you.
4. Using the T-10 driver, remove the eleven flat-top screws (1) from the inner
5. Using the T-10 driver, remove the screw (3) from the rear panel.
6. Remove the inner bottom cover.

Replacement Procedure

1. Reverse the order of the removal procedures.
2. Recommended torque for all screws is 9 in–lbs.
172
Figure 3-5
for this procedure.
on page
168.
bottom cover (2).