Running Self Test

Self Test is a series of internal tests of signal generator functions. If this test
fails, refer to
Use the following procedure to run self test:
1. Disconnect all external cables, including GPIB, LAN, and USB cables.
2. Preset the signal generator: Press Preset > Utility > Instrument Info > Sel f
3. Press Run Complete Sel f Test.
248
"Self Test Failure" on page 249
Test.
The following message appears:
• The current status of the self-test is: Incomplete.
Not all tests have been run.
An activity bar displays on the screen indicating the test progress.
If you press Abort while self-test is running, the following message displays:
The current status of the self-test is: Incomplete.
Not all tests have been run.
When self-test completes, one of the following messages displays:
The current status of the self-test is: Passed
The current status of the self-test is: Failure. One
or more tests have failed. System diagnostics
indicate this test as the root failure: xxx
for further instructions.
It takes about 5 minutes to run the self
test.
Opens a table in which you can select
tests and view details of those test.