Assembly Replacement
A8 Floating BNC Bypass (N5173B/83B only)
A8 Floating BNC Bypass (N5173B/83B only)
Tools Required
— 5/8" hex-nut driver
— needle-nose pliers
Removal Procedure
Refer to
1. Disconnect the power cord.
2. Remove the inner cover and outer top cover from the signal generator.
3. Disconnect the flexible cable from the Micro Deck at the location (1)
4. Using the 5/8" hex-nut driver, remove the nut and washer securing the A8
5. Remove the A8 Floating BNC Bypass.
Replacement Procedure
1. Reverse the order of the removal procedure.
2. Torque the nut to 21 in–lbs.
3. Perform the post-repair performance tests that pertain to this
Keysight X-Series Signal Generators Service Guide
Figure 3-38
for this procedure.
Refer to
"Outer Cover" on page
shown.
Floating BNC Bypass to the rear panel.
replacement procedure, see
Operational Performance
168,
"Inner Top Cover" on page
Chapter 4, "Post-Repair Procedures and
Verification".
170.
225