Rf Connector (N5173B/83B Only - Standard); Tools Required; Removal Procedure; Replacement Procedure - Keysight N5171B EXG Service Manual

X-series signal generators
Hide thumbs Also See for N5171B EXG:
Table of Contents

Advertisement

Assembly Replacement

RF Connector (N5173B/83B only - Standard)

RF Connector (N5173B/83B only - Standard)

Tools Required

— 5/16" open-ended wrench

Removal Procedure

Refer to
1. Disconnect the power cord.
2. Remove the covers from the signal generator. Refer to
3. Remove the front-panel from the signal generator. Refer to
4. Using a 5/16" open-ended wrench, disconnect the following semi-rigid
5. Remove the RF Connector.

Replacement Procedure

1. Reverse the order of the removal procedure.
2. Torque screw to 9 in–lbs.
3. Torque the new RF connector to 75 in–lbs.
4. Perform the post-repair performance tests that pertain to this
Keysight X-Series Signal Generators Service Guide
Figure 3-34
for this procedure.
page
168,
"Inner Top Cover" on page
page
172.
on page
174.
cable according to the signal generator you have.
520 with a 3.5mm RF connector: W10 from the RF connector (1).
520 with a Type-N RF connector: W11 from the RF connector (1).
532/540: W9 from the RF Connector (1).
replacement procedure, see
Operational Performance
170, and
"Inner Bottom Cover" on
Chapter 4, "Post-Repair Procedures and
Verification".
"Outer Cover" on
"Front-Panel"
217

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

N5183b mxgN5182b mxgN5172b exgN5173b exgN5181b mxg

Table of Contents