Keysight N5171B EXG Service Manual page 38

X-series signal generators
Hide thumbs Also See for N5171B EXG:
Table of Contents

Advertisement

Troubleshooting
Running Self Test
Table 1-1
Sel f Test Failure Number
100 Digital Self Tests
101
102
103
104
105
200 Reference Self Tests
201
202 - 208
300 Synthesizer Self Tests
301
302 - 320
400 Output Self Tests
401
402 - 413
500 Modulation Self Tests
501 - 506
507 Pulse Mod
600 Attenuator Self Tests
601
602 - 603
700 BB Digital Self Tests
701
702 to 708 (except 704)
704
800 BB Analog Baseband Self Tests
801 - 804
Keysight X-Series Signal Generators Service Guide
Self Tests
Replace the A3 RF Assembly
Go to
"Troubleshooting Self Test 102" on page 33
Go to
"A1 Power Supply Troubleshooting" on page 45
Replace the A3 RF Assembly
Replace the A3 RF Assembly
Go to
"A1 Power Supply Troubleshooting" on page 45
Replace the A3 RF Assembly
Go to
"A1 Power Supply Troubleshooting" on page 45
Replace the A3 RF Assembly
Go to
"A1 Power Supply Troubleshooting" on page 45
Replace the A3 RF Assembly
Error 405 for N5183B, go to
GHz (N5173B/83B)" on page 96
Replace the A3 RF Assembly
Go to
"Pulse Modulation Problems (N5173B/83B)" on
page 100
Go to
"A1 Power Supply Troubleshooting" on page 45
Replace the A3 RF Assembly
Go to
"A1 Power Supply Troubleshooting" on page 45
Replace the A2 Vector Module
Go to
"Troubleshooting Self Test 704" on page 33
Replace the A2 Vector Module
Resolution
"Checking ALC Operation < 3.2
31

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

N5183b mxgN5182b mxgN5172b exgN5173b exgN5181b mxg

Table of Contents