Post-Repair Procedures and Operational Performance Verification
Post Repair Procedures
Adjustments (N5173B/83B Only)
To ensure the signal generator is meeting all published specifications,
4-3
been repaired or replaced.
Download the service software from
http://www.Keysight.com/find/calibrationsoftware
or you can purchase the software on a CDROM (a license is required to run the
calibration software).
Table 4-3
Ad justments
MW Multiplier Pre-level Calibration
AM Calibration
Limited Absolute Power Calibration
ALC Mod Driver
244
lists the adjustments that must be performed after a specific assembly has
Adjustments
a
a. Assembly requiring adjustment is determined by the type
of failure (i.e. low band (A3 RF Assembly) or high band (A7
MW Assembly)).
Ad justed Assembly
X
X
X
X
X
X
Keysight X-Series Signal Generators Service Guide
Table