Quick Start
Verify Operation of the Signal Generator
Verify Operation of the Signal Generator
To verify the operation of the M9384B VXG, run a Self Test and Internal
Alignments.
Run a Self Test
Refer to the
Data Sheet
for a description of the front and rear panel
connectors.
1. Launch the M9384B application.
2. Select the Gear icon on the top right corner, and then select System.
3. Select the Diagnostics tab, then Run Self Test.
The self test takes 5 to 10 minutes to complete.