Assembly Replacement

Power Switch

Power Switch

Tools Required

— T-10 driver

Removal Procedure

Refer to
1. Disconnect the power cord.
2. Remove the covers from the signal generator. Refer to
3. Remove the front-panel from the signal generator. Refer to
4. Remove the A6 Front Panel Interface Scan assembly from the signal
5. Remove the LCD Display from the signal generator. Refer to
6. Using the T-10 driver, remove the two screws (1) that secure the Power
7. Remove the Power Switch.

Replacement Procedure

1. Reverse the order of the removal procedure.
2. Recommended torque for all screws is 9 in–lbs.
3. Perform the post-repair performance tests that pertain to this
206
Figure 3-29
for this procedure.
page
168,
"Inner Top Cover" on page
page
172.
on page
174.
generator. Refer to
"A6 Front Panel Interface Scan Assembly" on page
on page
201.
Switch (2) to the front-panel frame.
replacement procedure, see
Operational Performance
170, and
"Inner Bottom Cover" on
Chapter 4, "Post-Repair Procedures and
Verification".