Save As Dialog - Beckman Coulter Optima L-XP Instructions For Use Manual

Preparative ultracentrifuges l-xp series
Hide thumbs Also See for Optima L-XP:
Table of Contents

Advertisement

Operation
Programmed Operation
6
To save the program for later recall, press
use the on-screen keyboard (see
Figure 2.16 Save As Dialog
NOTE
If you do not save the program under a new filename, it will be overwritten when a new step
program is run.
7
(
Press
specific filename, the program is automatically saved to a temporary file called
LastStepProgram.spr. (Or press
8
(
Press
then press
or
(
Press
When all required parameters are entered, press
terminated at any time by pressing
2-18
)
to transfer the run parameters to the Settings window; if not saved to a
Download
(
)
to enter chamber precool or preheat temperature. Select temperature,
Precool Heat
(
)
. The vacuum will start automatically.
Start Temp
)
on the keypad to start or stop the vacuum pump.
Vacuum
(
)
. The Save As dialog box opens, and you can
Save
Figure
2.16) to specify a filename.
)
to close the program window without downloading it.)
Close
(
)
, then
Enter
(
)
on the keypad.
Stop
(
)
. A run may be
Start
PN LXP-IM-7AB