Bio-Safety Level 3 Installation; Electrical Requirements; Dimensions Of The Optima L-Xp Ultracentrifuge - Beckman Coulter Optima L-XP Instructions For Use Manual

Preparative ultracentrifuges l-xp series
Hide thumbs Also See for Optima L-XP:
Table of Contents

Advertisement

Preinstallation Requirements

Bio-Safety Level 3 Installation

The instrument will operate within specifications in a laboratory with ambient temperatures
ranging from 10 to 40°C.
Figure A.1 Dimensions of the Optima L-XP Ultracentrifuge
cm
in.
15.2
6
67.3
26.5
Bio-Safety Level 3 Installation
For laboratories with epoxy aggregate floors, such as BSL-3 labs, a non-invasive installation kit
(PN 393862) is available. The kit, which consists of a high-strength adhesive tape, is CSA certified for
use on epoxyed aggregate floors only.

Electrical Requirements

Instrument Rating:
Power Line Range:
To reduce the risk of electrical shock, this equipment uses a three-wire electrical cord (1.8 m; 6 ft)
and plug (see
instrument is supplied with an unterminated cord, a plug that meets local electrical and safety
requirements must be supplied. (Contact your local Beckman Coulter office for specific information
regarding these requirements.) See
matching wall outlet receptacle is properly wired and earth-grounded.
A-2
5.1
2
220 to 240 VAC, 50 Hz, 20 A
200 to 240 VAC, 50/60 Hz, 20 A
180 to 264 VAC, 60 or 50 Hz (single-phase), 30 A
Figure
A.2) to connect the equipment to earth ground. In regions where the
Table A.1
94.0
5.1
37
2
for the required wire connections. Make sure that the
PN LXP-IM-7AB