Edp Command Calibration - Rice Lake IQ PLUS 710 Installation Manual

Digital weight indicator
Hide thumbs Also See for IQ PLUS 710:
Table of Contents

Advertisement

Linearization is optional: if you choose not to
perform
linearization,
parameter; if linearization values have
previously been entered, these values are reset
to zero during calibration. To perform
linearization, follow the procedure below:
With
displayed, Press
WLIN
Þrst linearization point (
weights on the scale and press
Use the numeric keypad to enter the actual
test weight value, then press
calibrate. The indicator displays
calibration is in progress. When complete, the
A/D count for the linear calibration is
displayed. Press
calibration value and go to the next prompt
(
).
PT-> 2
Repeat for up to Þve linearization points. To
exit the linearization parameters, press
return to WLIN.
4.2

EDP Command Calibration

To calibrate the indicator using EDP commands, the
indicator EDP port must be connected to a terminal or
personal computer. See Section 2.3.2 on page 6 for
EDP port pin assignments; see Section 5.0 on page 40
for more information about using EDP commands.
Once the indicator is connected to the sending device,
do the following:
1. Place the indicator in setup mode (display
reads
) and remove all weight from
CONFIG
the scale platform. If your test weights require
hooks or chains, place the hooks or chains on
the scale for zero calibration.
2. Send the WZERO EDP command to calibrate
zero. The indicator displays
calibration is in progress.
3. Place test weights on the scale and use the
WVAL command to enter the test weight
value in the following format:
WVAL=nnnnnn<CR>
4. Send the WSPAN EDP command to calibrate
span. The indicator displays
calibration is in progress.
skip
the
WLIN
to go to the
PT-> 1
). Place test
or
ENTER
ENTER
while
*CAL*
again to save the
ENTER
while
*CAL*
while
*CAL*
7. The optional rezero function is used to
remove a calibration offset when hooks or
chains are used to hang the test weights.
¥
If no other apparatus was used to hang the
test weights during calibration, remove
the test weights and press
the CALIBR menu.
¥
If hooks or chains were used during
.
calibration, remove these and the test
weights from the scale. With all weight
to
removed, press
scale. This function adjusts the zero and
span calibration values. The indicator
displays
calibrations are adjusted. When complete,
the adjusted A/D count for the zero
calibration is displayed. Press
save the value, then press
to
the CALIBR menu.
8. Press
until the display reads
press
to exit setup mode.
5. Up to Þve linearization points can be
c a l i b r a t e d b e t w e e n t h e z e r o a n d s p a n
c ali b r at i o n va lu es . Us e t h e f ol low in g
commands to set and calibrate a single
linearization point:
WLIN.V1=nnnnn<CR>
WLIN.C1<CR>
The WLIN.V1 command sets the test weight
value (nnnnn) for linearization point 1. The
WLIN.C1 command calibrates the point.
Repeat using the WLIN.Vx and WLIN.Cx
commands
linearization points.
6. To remove an offset value, clear all weight
from the scale, including hooks or chains used
to hang test weights, then send the REZERO
EDP command. The indicator displays
while the zero and span calibrations are
adjusted.
7. Send the KUPARROW or KEXIT EDP
command to exit setup mode.
to return to
to rezero the
ENTER
while the zero and span
*CAL*
ENTER
to return to
CONFIG
as
required
for
additional
Calibration
to
, then
*CAL*
37