Kind
Description
Batch
Continuous
Ö
CONCUR
Allows a digital output to remain active over a specified portion of the batch
sequence. Two types of concur setpoints can be configured:
The digital output associated with this setpoint becomes
Type 1 (VALUE=0):
active when the START setpoint becomes the current batch step and remains
active until the END setpoint becomes the current batch step.
If a non-zero value is specified for the VALUE parameter, that
Type 2 (VALUE > 0):
value represents the timer, in tenths of a second, for this setpoint. The digital
output associated with this setpoint becomes active when the START setpoint
becomes the current batch step and remains active until the timer expires.
NOTE: If more than one concurrent setpoint is configured, each must be
assigned to a different digital output.
Table 8-1. Setpoint Kinds (Continued)
56
IQ plus 710 Installation Manual