1523, 1524 Reference Thermometer
Manual Calibration
Table 25 Test Equipment Specifications
Classification
Voltage Source
4 wire Reference Resistors
4 wire cable
Lemo Connector
Table 26 Standard Reference Resistor Specification
Reference
Resistor
Uncertainty
Resistance
(ppm)
0
--
25
28
75
15.73
100
14
200
11.5
400
10.25
10000
35
40000
26.25
100000
74
300000
70
500000
70
Table 27 Standard Voltage Reference Specification
7.7 Manual Calibration
7.7.1
General
This procedure outlines a manual process that closely follows the way the Reference
Thermometer was calibrated at the factory. The user should adjust their calibration
process to meet the needs of their facility and process. The following flow charts sum-
marizes the process.
78
Shop for Fluke products online at:
Minimum Use
± 6ppm ± 0.6uV
See Total Uncertainty column
Table X
Low resistance, Low EMF
6 pin
U
(k=2)
S1
Reference
Resistor
Uncertainty (Ω)
0.00004
0.0007
0.00117975
0.0014
0.0023
0.0041
0.35
1.05
7
21
35
Reference Voltage Total
Voltage mV
Uncertainty (mV)
–10
0.00125
0
0.00110
25
0.00145
75
0.00210
MyFlukeStore
www.
Recommended Equipment
Fluke 5700
Vishay VHP100 Resistors
N/A
LEMO PAG.M0.6GL.AC39G
U
(k=2)
S2
TCR
Uncertainty
TCR
(ppm)
Uncertainty (Ω)
--
--
0.3
0.0000075
0.3
0.0000225
0.3
0.00003
0.3
0.00006
0.3
0.00012
0.3
0.003
0.3
0.012
0.3
0.03
0.3
0.09
0.3
0.15
U
(k=2)
T
Total
Total
Uncertainty
Uncertainty (Ω)
(ppm)
0.000540
--
0.000700
28.0016
0.001180
15.7329
0.001400
14.0032
0.002301
11.5039
0.0041
10.2544
0.3500
35.0013
1.0501
26.2517
7.0001
70.0006
21.0002
70.0006
35.0003
70.0006
1.877.766.5412
.com