1523, 1524 Reference Thermometer
Serial Commands - Alphabetic Listing
The bit assignments are as follows:
bits 1 to 7:
bits 9 to 15:
STATus:MEASure:CONDition?
Return the Measurement Condition Register, indicating whether or not measuring is
active.
The bit assignments are the same as for STATus:MEASure?
STATus:MEASure:ENABle?
Return the Measurement Event Enable Register.
The bit assignments are the same as for STATus:MEASure?.
STATus:MEASure:ENABle <num>
Set the Measurement Event Enable Register.
The <num> parameter specifies the event enable bits which are to be set, 0 to 65535
decimal. The bit assignments are the same as for STATus:MEASure?
STATus:QUEStionable:EVENt?
Read and clear the Questionable Event Register, indicating any measurement
warnings.
The bit assignments are as follows:
68
Shop for Fluke products online at:
bit 0:
probe 1 new measurement
not used
bit 8:
probe 2 new measurement
not used
Example: STAT:MEAS?
Response: 257
Example: STAT:MEAS:COND?
Response: 257
Example: STAT:MEAS:ENAB?
Response: 0
Example: STAT:MEAS:ENAB 257
bit 0:
probe 1 not inserted
bit 1:
probe 1 measurement under range
bit 2:
probe 1 measurement over range
bit 3:
probe 1 EEPROM fault (bad checksum or
version number is not supported)